Technical Features
Introduction of KTLDE Electromagnetic Flowmeter
Measuring Principle
How to Select Meter Model Correctly
How to Select Electrode Material
How to Select Liner Material
How to Select Meter Protection Grade & Extra-function
How to Select Meter Installing Location
Installing Site
Wiring
Pattern of Remote Type and Dimensions of Converter
Structure of Fixed Inserted Electromagnetic Flowmeter & Dimensions
Structure of Adjustable Inserted Electromagnetic Flowmeter & Dimensions
Meter Dimensions & Weight
Introduction of Multi Electrodes
Blade Electrode Maintenance and Cleaning
Sanitary Flowmeter Application and Selection
Model Selection
Meter Specification
Pressure Lose of the Diverging and Converging Pipe equipped
KTLDE Electromagnetic Flowmeter
 
 Pay Special Attention to Installing Site
 
¢ Axial line of the measuring electrode must be horizontal
  

¢ Meter pipe must be filled with the measured fluid fully
¢ Specify the minimum length of the straight pipe 5D (D=inside diameter of the flowmeter) for upstream, 3D for downstream. In order to install and remove flowmeter conveniently, suggest a flexing pipe be used behind the flowmeter




¢ Flowing direction must be conformed to the indicating arrow on the flowmeter
¢ Negative pressure in pipe line will destroy the liner of flowmeter
¢ No strong magnetic field and radio around the meter
¢ To install and maintain meter conveniently, there should be enough space around the meter
¢ A fixed bracket be equipped at both side of the flowmeter if the pipe vibration existence
¢ The distance of the meter to mixing point should be 30D (D=inside diameter of the flowmeter) if the measured fluid is mixed
¢ To convenient cleaning and maintenance, a by-pass pipe should be equipped, as well as ensure the condition of straight pipe 5D for upstream and 3D for downstream




¢ When install meter, tight the bolts of the flange with a average torque force to prevent liner from broken, using special torque wrench

¢ ? Meter be installed in the lower pipe or upward vertical pipe, avoid installing at the highest point in pipeline or in the downward vertical pipe
¢ Install where pipe is rising
¢ For draining pipe, meter be installed at the lower point
¢ If the differential fall of the pipe is >5m, a draining valve for gas be installed in the downstream of the meter
¢ Control valve and cutoff valve be installed in the downstream of the meter instead of in the upstream
¢ The meter must be installed at the outlet of the pump, but not at the inlet
¢ Meter installed in well

1. Inlet      2. Overflow-tube
3. Inlet grid   4. Cleaning hole
5. Flowmeter  6. Duct
7. Outlet      8. Draining valve
9. Draining valve for dirties



 